10 troubleshooting, Eppendorf reference® – part a – chapter 10 – Eppendorf Reference User Manual

Page 20

Advertising
background image

19

Error

Cause

Solution

Droplets on
the inner wall
of the pipette
tip.

– Uneven wetting of the

plastic wall.

– A pipette tip with poor

wetting properties has
been used.

– Attach a new pipette tip.

– Use an original eppendorf

tip.

Pipette is
dripping
and/or the
volume
pipetted is
incorrect.

– The tip is loose.
– A poorly fitting pipette tip

has been used.

– Liquid with a high vapor

pressure has been pipetted.

– Tip has been taken out of

the liquid too quickly.

The pipette is dripping
because:
– Piston is contaminated.

– Piston is damaged.

– Seals are damaged.

– Tip cone loose.

– Press the tip firmly in place.
– Use an original eppendorf

tip.

– In this case, we recommend

pipetting using a positive-
displacement pipette.

– Remove the tip slowly from

the liquid.

– Clean and lightly lubricate

the piston.

– Replace the piston and seal

(see Part B, Maintenance).

– Replace all seals

(see Part B, Maintenance).

– Lightly tighten tip cone with

wrench (see Part B,
Maintenance).
Exchange, where
necessary.

10

Troubleshooting

Eppendorf Reference® – Part A – Chapter 10