Control button
jams, moves
erratically.

– Piston contaminated.

– Seals contaminated.

– Penetration of solvent

vapors.

– Clean piston and lubricate

lightly.

– Disassemble pipette. Clean

all seals and exchange
where necessary
(see Part B, Maintenance).

– Unscrew pipette at the

central junction and
ventilate.
Clean piston and lubricate
lightly.

Pipette
blocked,
too little liquid is
aspirated.

– Liquid has penetrated the tip

cone and dried.

For 25 to 500 μL pipettes:
– The filling tube in the nose

cone is blocked.

– Unscrew pipette at the

central junction, rinse lower
part first with warm water,
then with distilled water and
allow to dry.

Or:
– Disassemble pipette.

Replace ejector seal in the
tip cone (see Part B, Mainte-
nance).

For 25 to 50 μL:
– Replace tip cone.

For 100 to 500 μL:
– Replace the filling tube in

the tip cone (see Part B,
Maintenance).

Error

Cause

Solution