Step 1: Point the sensor at a target that is approximately at the
maximum measurement range. Adjust the MulƟ plier value so
that the sensor distance reading matches the actual measured
distance to the target.
Distance CalibraƟ on
Use a wall or other large Ň at target for calibraƟ on at longer ranges.
Step 3: Repeat steps 1 and 2 unƟ l the reading at each endpoint
matches the actual measured distance. The accuracy at both end-
point should improve with each repeƟ Ɵ on of steps 1 and 2.
measure
measure
Step 2: Point the sensor at a target that is approximately at the
minimum measurement range. Adjust the Oī set value so that the
sensor distance reading matches the actual measured distance to
the target.