Installation
- Unpacking -
When unpacking the instrument, exercise care not to subject the instrument to
mechanical shock. After unpacking, visually inspect the instrument for
damage.
- Environment -
The FLE series sensors should be installed in an area which meets the following
conditions:
1. Non-hazardous area.
2. The medium temperature does not exceed -14
0
F to 185
0
F (-10
0
C to 85
0
C).
3. Locate the sensor away from strong magnetic fields such as those produced
by motors, transformers, solenoid valves, etc.
4. The medium is free from metallic substances and other foreign matter.
5. No corrosive gases such as NH
3
, SO
2
, Cl
2
, etc.
6. No excessive vibration
7. Ample space for maintenance and inspection.