Program review – Carbolite 302P4 User Manual
Page 15

Programming
15
MC22-GB-V1.00
Time Held (Dwell)
After confirming selection, the display will then show, tH.1 lower display and the value in the
upper display.
Use the Up
or Down
to select a value for time held (dwell).
Press
to confirm
Set Point 2 onwards
After confirming selection, the display will then show, SP.2 lower display and the value in the
upper display.
Use the Up
or Down
to select a value for set point 1, or reduce the value to below the
minimum to select End, if no further segments are required
Press
to confirm.
This Process is repeated for each of the 4 segments of the program.
Repeat Program
After confirming selection, of the last parameter in the program, the display will then show,
rPt. lower display and the value in the upper display.
Use the Up
or Down
to select how many times the program requires repeating
Press
to confirm
Cycle End Temperature
After confirming selection, of the last parameter in the program, the display will then show,
CEt lower display and the value in the upper display.
Use the Up
or Down
to select the temperature at which the program cycle will end.
Note: Below Zero the cycle end temperature will display off, when set at this value the
program will display EnD after last dwell period.
Press
to confirm