5.1

Customer Calibration

The 302P controller is calibrated for life at manufacture, there may however be sensor or
other system errors, which affect the accuracy of the measured temperature. Customer
calibration can be used to compensate for these errors.

Ensure the controller is in the Enhanced mode OP.E

1 Point Calibration CL.St = C.CL1

1 point calibration uses one offset value across the temperature range.

Clear any previous coefficient adjustment.

Press function

until CL.St is shown in the lower display


Press the Up

or Down

to C.CL1 in the upper display


Press function



to confirm selection and display OFSt


Press the Up

or Down

to enter the offset value required across the temperature range.

This can be a negative or positive value.

Press function



to confirm selection.




2 Point Calibration CL.St = C.CL2

2 point calibration uses two offset values at two corresponding temperatures; this changes
the calibration linearly as the temperature increases or decreases.

Clear any previous coefficient adjustment.

Press function

until CL.St is shown in the lower display


Press the Up

or Down

to C.CL2 in the upper display


Press function



until CAL.L is displayed


Press the Up

or Down

to enter the lower calibration temperature to which the offset is to

be applied. Example. 300°C.

Press function



until OFS.L is displayed


Press the Up

or Down

to enter the offset to be applied to the lower calibration

temperature. Example: If at the calibrated temperature of 300°C the controller is reading
296.7, the offset will be 3.3°C (300-296.7 = 3.3)