Operation – INFICON TGF10 Tracer Gas Filler User Manual

Page 7

Advertising
background image

EN

INFICON – User’s manual TGF10 5

1.3. Operation

Leak Detection Sequence:

1) Press the start button.

a) The “Evac” diode comes on and the test object is evacuated.

Pre fill evacuating time is set and adjusted by a timer in Sensistor

ISH2000.

b) The “Fill” diode comes on and the “Evac” diode goes off.

Note: The “Evac” diode goes off shortly after the “Fill” diode
has come on.
Test object is now filled with tracer gas. Filling pressure is

controlled by the gas bottle regulator.









2) Search and locate leak(s).











3) Press stop button.

a) The “Evac” diode comes on and the test object is evacuated.
b) After test evacuation time is set by a timer in Sensistor

ISH2000.

c) Evacuation stops and Test object is refilled with air to

atmospheric pressure