INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 21

Binary Protocol
1-19
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
32
33
GetRangeNV
SetRangeNV
Get set post amplifier range
No
unsigned char
auto: 0-0.4; 1-1.6; 2-6.4; 3-
25.6
manual: 4-0.4; 5-1.6; 6-6.4; 7-
25.6
34
35
GetR43
SetR43
Get / set 500GOhm => 15GOhm factor
No
float
36
37
GetCalFac
SetCalFac
Get / set calibration factor
Byte 0:
0: mass2 /vacuum
mode
1: mass3 /vacuum
mode
2: mass 4 /vacuum
mode
3: mass 2 /
sniff mode
4: mass 3 /
sniff mode
5: mass /
sniff mode
6: mass 2 /
auto leak test
7: mass 3 /
auto leak test
8: mass 4 /
auto leak test
9: mass 2 /
commander
10: mass 3 /
commander
11: mass 4 /
commander
float
38
GetMinSinceStart
Get time since power on
No
unsigned long integer [min]
39
GetZustArange
Get Autoranging state
No
unsigned char
40
41
GetMass
SetMass
Get / set mass (0/1/2 for mass 2/3/4)
No
unsigned char
42
43
GetMFAE
SetMFAE
Get / set anode voltage
Byte 0:
Mass (0/1/2 for
mass 2/3/4)
unsigned integer [V]
44
GetMSCtrl
Get MSV state
No
2 Bytes
Byte 0: Nominal condition (0-
off,
1-Saving mode, 2-Normal
mode)
Byte 1: Actual condition (0-off,
1-Saving mode, 2-Normal
mode)
45
SetMSCtrl
Set MSV state
No
unsigned char
(0-off, 1-Saving mode,
2-Normal mode)
No
Name
Description
Parameter
Data format