2 interface, Connecting, Cable – INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 4: 2interface connecting cable

1-2
Interface Connecting Cable
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
2
Interface Connecting Cable
The interface (RS232) is wired as data communication equipment (DCE). The
connection is made via a 9-way sub-D socket at the leak detector. The signals are
assigned as follows:
The other pins are not used.
The levels on the RS 232 interface are defined as follows:
RS232 Connecting Cable
A standard RS232 cable can be used (straight-through connecting cable, RxD and
TxD not crossed).
The RS232 hardware handshake must be switched off (in RS232 control program
written by the user).
If switching off of the hardware handshake is not possible, an RS232 connecting
cable wired as follows may also be used:
Pin Name
Signal
2
RXD
Receive data (Modul1000
→ PC)
3 TXD
Transmit
data
(PC
→ Modul1000)
5 GND
Reference
ground
Level Low
(L)
High
(H)
Voltage range
-3V … - 25V
3V … 25V
Logic state
logical 1
logical 0
Level designation
Mark
Space
Fig. 1-1 RS232 Connecting Cable