INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 22

1-20
Binary Protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
46
47
GetMSKat
SetMSKat
Get / set active filament
No
unsigned char
(0=Cathode 1, 1=Cathode 2)
48
49
GetMSConfig
SetMSConfig
Get / set MS configuration
No
2 Byte
Byte 0: MFAE
Command variable
Anode potential externally
Byte 1: METS Suppressor test
function
50
51
GetZero
SetZero
Get / set Zero On/Off
No
uchar
0=off / 1=on
52
Start
Start
No
No
53
Stop
Stop
No
No
54
GetCal
Get calibration state
No
uchar
0=idle
1=evacuation
2=wait_stable_open
3=auto_tune
4=auto_tune_restart
5=wait_close
6=wait_stable_close
7=wait_okay
55
Cal
Start calibration
No
No
56
57
GetTrigger
SetTrigger
Get / set trigger
Byte 0:
1...3 for Trigger
1...3
Byte 1:
Unit: 0-mbar*l/s,
1-Pa*m³/s, .
2-atmcc/s,
3-Torrl/s;
at Sniff
additionally: 4-
ppm, 5-g/a
float
58
59
GetOpMode
SetOpMode
Get set operation mode
No
uchar
0=Commander
1=Auto Leak Test
2=Vacuum
3=Sniff
60
61
GetGasballast
SetGasballast
Get / set Gasballast
Byte 0:
0 - close gasballast
/ purge valve
1- open gasballast
/ purge valve in
standby
No
Name
Description
Parameter
Data format