To edit film parameters:

Press the fixed “Menu” button to view the Menu screen.


Press “Review Film” on the touchscreen to view the film parameter edit screen.


Select "GO2 Flm" on the touchscreen, and enter the number of the film you want to review/edit
(select the number using the touchscreen keypad, then select "Enter".)



When the film numbers next to "Edit" and "Active" at the top of the screen match, you are
editing the active film.


Note: Changes made to the active film will change a running process.

To Start:

From any Status screen, press the fixed “Start” button.


The word “Ready” should be displayed in the upper left corner of the screen.
Press the start button again to start deposition.