4 - 2

PN

07

4-

60

9-

P1

A

Cool Drawer Single and Dual Sensors Operating Manual

A very useful tool for rapidly evaluating the cause of a persistent "Crystal Fail" is
the test crystal, which is included with each non-modelock oscillator package, this
tool utilizes a packaged crystal at 5.5 MHz and a connector that allows the direct
connection to BNC cables.
The test crystal provides a known "good" monitor crystal that provides a fast means
of isolating sensor problems.

CAUTION

The test crystal is designed as a diagnostic tool, and is
not intended for use in vacuum.

4.1.2 Diagnostic Procedures

The following diagnostic procedures employ the test crystal and DMM to analyze a
constant Crystal Fail message. The symptom is a Crystal Fail message that is
displayed by the deposition controller even after the monitor crystal has been
replaced with a new “good” monitor crystal.

4.1.2.1 Measurement System Diagnostic Procedure

1

Remove the 15.2 cm (6 in.) BNC cable from the feedthrough.

2

Connect the test crystal to the BNC cable.

If the Crystal Fail message disappears after approximately five seconds,
the measurement system is working properly. Reinstall the BNC cable to
the feedthrough. Go to

section 4.1.2.2

.

If the Crystal Fail message remains, continue at step 3.

3

Disconnect the BNC cable from the oscillator and from the test crystal.

4

Visually inspect the BNC cable to verify that the center pins are seated properly.

5

Use a DMM to verify the electrical connections on the BNC cable, refer to

section 4.1.1

.

There must be continuity between the center pins.

There must be isolation between the center pins and the connector shield.

There must be continuity between the connector shields.

Replace the BNC cable if it is defective and repeat step 2 of this procedure.

6

If the BNC cable is not defective, reconnect it to the oscillator and the test
crystal. If the Crystal Fail message remains, contact INFICON (refer to

section

1.3 on page 1-3

).