PLO-10 PHASE LOCK OSCILLATOR

UNDERSTANDING AND SETTING UP THE INFICON PLO-10

2-2

Connect a frequency counter to the Frequency Output.
Connect a voltmeter to the Conductance output. The center conductor on the BNC

connector is positive with respect to ground.
Connect the crystal holder, with a crystal installed, to the PLO by means of the 12-inch

coax cable.
Plug the wall mount power supply into the wall and plug the power plug into the PLO-10.
Refer to Figure 2, Figure 3, and Figure 4 for a complete system connections.
The green, Lock, LED should come on, the frequency should indicate the correct crystal

frequency and the voltmeter should indicate something between 5 millivolts and 4 volts.
Check the capacitance cancellation by pressing and holding the Reset switch. The green,

Lock LED should light. Keeping the Reset switch pressed, adjust the fine capacitance

trimmer clockwise by about 5 degrees. The yellow, Reset LED should flash. Back the

trimmer counterclockwise to the point where the Reset LED just stops flashing. The

capacitance cancellation should be checked and readjusted every time the environment of

the crystal and holder is changed. For example, if the crystal and holder are moved from

air to liquid or liquid to air, the capacitance cancellation should be checked and

readjusted.
Remove the crystal. The red, Unlock, LED should light. The green, Unlock, LED

should go off. The Sweep Rate LED should not flash. If the Sweep Rate LED flashes

the capacitance is under compensated.