Crystal cleaning, General cleaning, Organic (hydrocarbon contaminants) – INFICON PLO-10i Phase Lock Oscillator User Manual

Page 42: Lipid vesicles on sio2 surfaces, Crystal cleaning -8, General cleaning -8, Organic (hydrocarbon contaminants) -8, Lipid vesicles on sio, Surfaces -8

Advertising
background image

PLO-10 PHASE LOCK OSCILLATOR

CRYSTALS, HOLDERS AND FLOW CELL

7-8

dry the crystal. It is better to chase liquid off the crystal than to let it evaporate off the

crystal.

Never wipe the crystal – even soft, lint-free cloth will scratch the crystal.

7.2.1 CRYSTAL

CLEANING

The surface properties of the sensor crystal determine the interaction of sample material

with the surface. Therefore, the developments of proper procedures for cleaning are

required to obtain meaningful and reproducible measurements. This section provides the

basic information you need to develop a cleaning procedure suited to your sample/surface

preparation.

CAUTION

– When developing a cleaning procedure, always perform

a test run on a crystal before committing to a larger batch cleaning.

Follow the crystal handling guidelines throughout the cleaning process

to protect the crystal quality. Avoid using high pH cleaners since they

will etch the crystal surface.

7.2.1.1 General

Cleaning

For general purpose applications such as electrochemistry and liquid or viscoelastic film

experiments, it is usually sufficient to use ultrasonic cleaning method to clean the crystals

in a solution of non-basic detergent in deionized water. Immediately rinse liberally with

deionized water and dry in a gentle flow of filtered nitrogen gas.

7.2.1.2 Organic (hydrocarbon contaminants)

UV/ozone treatment

12

is a powerful tool for removing hydrocarbon impurities which

have been adsorbed from the ambient air. This method utilizes irradiation with ultra

violet light that breaks up the organics on the surface of the sample being cleaned. A

flow of air or a weak vacuum carries off the organics. This method does not affect the

quartz surface; it is low cost and is very efficient.
Oxygen plasma cleaning is another effective method that will remove organic matters. In

this method, the plasma reaction breaks up organic matters at the surface of the sample

being cleaned into smaller molecules and a vacuum pump removes them from the surface

of the sample.

7.2.1.3 Biomaterials (lipids, proteins and similar biomolecules)

Start by treating the crystal in an UV/ozone chamber for 10 minutes, then immerse it into a

1:1:5 solution of hydrogen peroxide (30%), ammonia (25%) and deionized water heated

to a temperature of about 75°C for 5 minutes

22

. Immediately rinse liberally with

deionized water and dry in a gentle flow of nitrogen gas. Immediately before

measurement, treat the crystal with UV/ozone for 10 minutes.

7.2.1.4 Lipid vesicles on SiO

2

surfaces

Treat the crystal in an UV/ozone chamber for 10 minutes, then immerse it into water with

2% of sodium dodecyl sulfate (SDS) at room temperature for 30 minutes

23

. Rinse

generously with deionized water and blow dry with filtered nitrogen gas. Immediately before

measurement, treat the crystal with UV/ozone for 10 minutes.

Advertising