PLO-10 PHASE LOCK OSCILLATOR

CRYSTALS, HOLDERS AND FLOW CELL

7-5

small mass changes over long periods of time.

7.1.8 CRYSTAL LIFE EXPECTANCY

It is difficult to predict the useful life of a crystal since it depends on many factors. Some

of these factors are

11

:

The quality of the quartz

The amount of deposited material

The stress generated in the crystal due to deposited material

The acoustic losses in the deposited material

The design of the oscillator circuitry

Other aspects that affect the crystal life include the type of the deposited material,

splitting of source material resulting in non- uniform films, film flakes that landed on the

crystal’s active area, and of course, physical damage to the crystal such as chipping,

cracking, or peeling of the electrode, etc.
In general, a sensor crystal can be used until its frequency drops below 50% of its

uncoated value. However, for the reasons stated above, crystal failures often occur well

before a 40% shift in frequency is reached.
The sensor crystals are considered expendable. However, a crystal may be reused up to

20 times on average in experiments that don’t physically alter the crystal electrode. In

experiments where a film is deposited, the crystal can be stripped using a chemical

etchant . Care must be taken so only the deposited material is stripped and not the crystal

electrodes. The amount of times that a crystal can be reused greatly depends on its

condition after each experiment or stripping. Needless to say, careful handling and

cleaning of the crystal is required to maximize its re-usability.
Noisy or erratic measurement indicates that the crystal is about to fail. It might even be

difficult to obtain a stable baseline. Spurious signals might become evident in

electrochemical QCM experiments. Visually, traces of consumption and wear can often

be seen on the crystal surface. Edges of the sensor crystal might become cracked and the

deposited film, even the electrode, starts to show scratches and tears.
The crystal motional resistance R does reflect the influence of deposited material on the

performance of a crystal. This resistance is associated with the damping of acoustic

waves by the electrodes, deposited materials, and the supporting structure. This

resistance increases as more material is being deposited onto the crystal

11

. This

resistance value can be used to determine when a crystal reaches a maximum loading.

7.1.9 TEMPERATURE

COEFFICIENT

The temperature coefficient of quartz crystals is normally specified in units of parts per

million per degree of temperature change. A one part per million change in frequency of

the sensing crystal corresponds to an indicated thickness change of approximately 7.4 Å

for a material with a density of 1.0 gm/cm³. For Aluminum with a density of 2.7 gm/cm³,

this is equivalent to approximately 2.7Å. This intrinsic dependence of resonance