Installation, Measurement signal vs. pressure, Gas type dependence – INFICON PSG512-S Pirani Standard Gauge User Manual

Page 2: Vacuum connection, P = 10, U = c + 1.286 × log, C × pressure reading

Advertising
background image

Internal volume

DN 16 ISO-KF
DN 16 CF-R
1/8" NPT
8 VCR

®

4 VCR

®

½"-Rohr
7/16-20 UNF

cm

3

cm

3

cm

3

cm

3

cm

3

cm

3

cm

3

≈1.5

≈1.5

≈2

≈2

≈2

≈2

≈1.5

DN 16 ISO-KF long tube
DN 16 CF-R

long tube

cm

3

cm

3

≈10

≈10

Admissible pressure

bar
(abs.)

10, limited to inert

gases

Admissible temperatures

Operation °C

+5 … +60

Vacuum connection

DN 16 ISO-KF
DN 16 CF-R
1/8" NPT
8 VCR

®

4 VCR

®

½"-Rohr
7/16-20 UNF

°C
°C
°C
°C
°C
°C
°C

80

2)

80

2)

80
80
80
80
80

in horizontal
mounting ori-
entation

Filament °C

110

Storage

°C

–20 … +65

Relative humidity

%

≤80 at temperatures up to
≤+31 °C, decreasing to 50
at +40 °C

Use

indoors only, altitude up to
2000 m NN

Mounting orientation

any

Protection category

IP40

Dimensions mm

70.

3

20.9

19

tube ½"

7/16-20 UNF

24

8 VCR

31.

5

41.

5

4 VCR

1/8" NPT

29

DN 16 ISO-KF

DN 16 CF-R

115

116

14

15

*

)

DN 16 ISO-KF

DN 16 CF-R

*

)

PSG51x-S: 20 mm

50

65

32.8



Weight

DN 16 ISO-KF
DN 16 CF-R
1/8" NPT
8 VCR

®

4 VCR

®

½"-Rohr
7/16-20 UNF

g
g
g
g
g
g
g

80
100
70
130
100
70
80

DN 16 ISO-KF long tube
DN 16 CF-R

long tube

g
g

130
140


2)

250 °C with long tube.

Measurement Signal vs. Pressure

1E+6

1E+4

1E+3

1E+2

1E+1

1E+0

1E–1

1E–2

1E–3

1E–4

0.0 0.5 1.0 1.5

2.5

3.5

4.5

5.5

6.5

7.5

8.5

9.5

10.3

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

mbar

Pa

1E+5

Torr

U

n

d

e

rr

ang

e

S

ens

o

r er

ror

Pressure p

[mbar]

Measurement signal U[V]

Over

ran

ge


p = 10

((U-c)/1.286)

U = c + 1.286 × log

10

p

valid in the range 5×10

-4

mbar <p< 1000 mbar

3.75×10

-4

Torr <p< 750 Torr

5×10

-2

Pa <p< 1×10

5

Pa

U p

c

U p

c

[V] [mbar] 6.143

[V] [micron] 2.448

[V] [µbar] 2.287

[V] [Pa] 3.572

[V] [Torr] 6.304

[V] [kPa] 7.429

[V] [mTorr] 2.448

where p pressure

U measurement signal
c constant (depending on pressure unit)


Gas Type Dependence

Pressure reading (gauge adjusted for air)

10

2

8

6
4

2

10

1

8
6
4

2

10

0

8

6
4

2

10

–1

8

6
4

2

10

–2

8
6
4

2

10

–3

10

–3

2

4

6

10

–2

2

4

6

10

–1

2

4

6

10

0

2

4

6

10

1

2

4

6

10

2

H

2

He

Ne

Xe

Kr

Ar

CO

2

Freon 12

Air

O

2

CO

N

2

p (mbar)

p

eff

(mbar)

Water vapor

Calibration factors for the pressure range below 1 mbar

p

eff

= C × pressure reading

Gas
type

Calibration

factor C

Gas

type Calibration

factor C

He
Ne

Ar
Kr

Xe

0.8
1.4
1.7
2.4
3.0

H

2

air, O

2

, CO, N

2

CO

2

water vapor

freon 12

0.5
1.0
0.9
0.5
0.7





Installation

Vacuum Connection

DANGER

DANGER: overpressure in the vacuum system
>1 bar
Injury caused by released parts and harm
caused by escaping process gases can result if
clamps are opened while the vacuum system is
pressurized.
Do not open any clamps while the vacuum sys-
tem is pressurized. Use the type of clamps which
are suited to overpressure.

DANGER

DANGER: overpressure in the vacuum system
>2.5 bar
KF connections with elastomer seals (e.g.
O-rings) cannot withstand such pressures. Pro-
cess media can thus leak and possibly damage
your health.
Use O-rings provided with an outer centering
ring.

DANGER

DANGER: protective ground
Incorrectly grounded products can be extremely
hazardous in the event of a fault.
The gauge must be electrically connected to the
grounded vacuum chamber. This connection
must conform to the requirements of a protective
connection according to EN 61010:
• CF, NPT, VCR

®

and UNF connections fulfill

this requirement.

• For gauges with a KF connection, use a con-

ductive metallic clamping ring.

• If a ½" tube is used, take appropriate meas-

ures for this requirement to be fulfilled.

Caution

Caution: vacuum component
Dirt and damages impair the function of the vac-
uum component.
When handling vacuum components, take ap-
propriate measures to ensure cleanliness and
prevent damages.

Caution

Caution: dirt sensitive area
Touching the product or parts thereof with bare
hands increases the desorption rate.
Always wear clean, lint-free gloves and use
clean tools when working in this area.

The gauge may be mounted in any orientation. To

keep condensates and particles from getting into the
measuring chamber preferably choose a horizontal
to upright position and possibly use a seal with a
centering ring and filter. If adjustment should be pos-
sible after the gauge has been installed, be sure to
install it so that the button can be accessed with a
pin (

→ "Adjusting the Gauge").

Remove the protective lid and install the product to the vac-
uum system.

Keep the protective lid.




t i na44e1- g

(2010-06)

Original: German tina44d1-g (2010-06)