Remove the protective lid and install the product to the
vacuum system.

Keep protective lid

Clamp

Seal with

centering ring

Power Connection

(HPG400)

The following information on the electrical connec-
tion as well as the wiring diagram apply to HPG400
only (

→  [1] and [2] for details on the electrical

connection and additional functions of HPG400-SD
and -SP).

Make sure the vacuum connection is properly made

(

→ "Vacuum Connection").

If no sensor cable is available, make one according to

the following diagram.

Electrical connection

Pin 2 Signal output (measuring signal) 0 … +10 V
Pin 5 Supply common, GND
Pin 8 Supply +24 VDC
Pin 10 Gauge identifcation
Pin 12 Signal common, GND
Pin 13 RS232C, TxD
Pin 14 RS232C, RxD
Pin 15 Shielding, housing, GND
Pins 1, 3, 4, 6, 7, 9 and 11
are not connected internally.

8

9

1

15

TxD

HPG400

56kΩ

RxD

Measuring signal

+Ub

13
14

2

12

8
5

15

+

1.25AT

Identification

10

RS232C

+

24 V

D-Sub, 15-pin

female

soldering side

Connect the sensor cable to the gauge.

Secure the cable connector with the lock screws.

Connect the sensor cable to the controller.



Operation

When the voltage is supplied, the measuring signal is avail-
able between pins 2 (+) and 12 (–) (Relationship Measuring
Signal – Pressure

→ "Technical Data" and  [1]).

HPG400-SD and -SP can also be operated via the corre-
sponding fieldbus interface (DeviceNet or Profibus) (

→  [1]

and [2]) for further details and functions).

Allow for a stabilizing time of

≈10 minutes. Once the gauge

has been switched on, permanently leave it on irrespective of
the pressure.

Gas Type Dependence

The measurement value is gas dependent. The displayed
reading applies to dry air, O

2

and N

2

. For other gases, it has

to be converted (

→ "Technical Data" and  [1]).

Measuring Range

The HPG400 covers the measuring range
2×10

-6

mbar ... 1000 mbar.

• The Pirani part permanently monitors the pressure.
• The hot cathode part (controlled by the Pirani) is only

switched on when the pressure drops below the change-
over threshold. The hot cathode will be ready for operation
after a few seconds heating time, when the <EMI ON>
lamp is lit.

• When the pressure rises above the changeover threshold,

the hot cathode is switched off and the <EMI ON> lamp
turns off.

In the upper pressure range, the Pirani reading and in the
lower pressure range, the hot cathode reading is output.

1000 mbar

2 ×

10

-6

mbar

Hot cathode HP

Pirani

5×10

-2

mbar ... 1 mbar user-

definable changeover threshold

Selecting the Changeover Threshold

The HPG400 has five user-definable changeover thresholds.
It is thus possible to prevent the changeover range from be-
ing situated within the process pressure range. The factory
setting of the threshold is 1 mbar. Another changeover
threshold can be selected via the <P

↔ HP> switch. Since

the contamination of the hot cathode part is reduced at low
pressures, the lowest possible changeover threshold should
be selected.

Switch position (<P

↔ HP>)

Changeover threshold

0 or 1

1 mbar (factory setting)

2 or 3

5×10

-1

mbar

4 or 5

2×10

-1

mbar

6 or 7

1×10

-1

mbar

8 or 9

5×10

-2

mbar

Since the switch position is only polled upon activa-

tion of the gauge, the changeover threshold should
be selected before the gauge is turned on.


Adjusting the Gauge

(HPG400)

The adjustment of HPG400-SD and -SP (

→  [1]

and [2]) is slightly different from the procedure for
HPG400, which is described below.

The gauge is factory calibrated. If used under different cli-
matic conditions, at extreme temperatures, through aging or
contamination and after exchanging the sensor, the char-
acteristic curve can be offset and readjustment can become
necessary. Only the Pirani element can be adjusted and only
at atmosphere.

Readjustment becomes necessary if
• at atmosphere the output voltage is <9.75 V or the display

reading is <atmosphere

• when venting the vacuum system, the output voltage

reaches 9.75 V before the measured pressure has
reached atmosphere (Gauges with display will show the
error "5" at atmosphere (Pirani sensor warning)).

(For more details

→  [1], "Maintenance, Repair").

Adjustment of the Pirani part under high vacuum
conditions:
The Pirani part is automatically adjusted by the hot cathode
part when the gauge is activated and the pressure range
1 … 3×10

-3

mbar is reached for the first time.

Adjustment of the Pirani part at atmospheric pressure:

Put the gauge into operation.

Operate the gauge for

≈10 minutes at atmos-

pheric pressure. If the gauge was operated
within the hot cathode range, a cooling-down
time of

≈30 minutes is to be expected (gauge

temperature = environmental temperature).

Adjust the gauge

HPG400 without display

353-520 353-522

HPG400 with display

353-521 353-523

Insert a pin (

≈ш1.3Ч50mm) through the opening

<FULL SCALE> and push the button inside

for at least 5 seconds.



≈10 s

Automatic adjustment in progress.

Adjustment completed.



Adjusting the Calibration Setting of the
Hot Cathode Part

The sensor is factory calibrated. The calibration setting of the
hot cathode range 0 … F is printed on the label. Correct this
value with the <Cal adj> switch to adjust the electronics to
the sensor. Before operating the gauge for the first time or
after replacing the sensor, check the calibration value setting
and adjust it if necessary.

Display

(HPG400 with part numbers 353-521 and 353-523)

Pressure reading

Pressure unit

Status