Deinstallation, Maintenance, repair, Returning the product – INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 4: Disposal, Further information, Ec declaration of conformity, High pressure / pirani gauge

Advertising
background image

Parameter Display

Description

Reading

0 … 1000

mbar, Torr, Pa

Pressure
unit

mbar, Torr, Pa Factory setting: mbar

Status (none)

E
A

ok
Fail IonG 5
Fail IonG 8

Fail IonG 9
no Signal

Pirani operation
Emission hot cathode on
Adjustment at atmospheric
pressure in progress
Normal operation, no error
Pirani sensor warning

1)

Hot cathode sensor
warning

1)

Pirani sensor error
Internal data connection
failure

1)

1)

Background illumination red (in normal operation green).



Deinstallation

DANGER

DANGER: contaminated parts
Contaminated parts can be detrimental to health
and environment.
Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant
regulations and take the necessary precautions
when handling contaminated parts.

Caution

Caution: vacuum component
Dirt and damages impair the function of the vac-
uum component.
When handling vacuum components, take ap-
propriate measures to ensure cleanliness and
prevent damages.

Caution

Caution: dirt sensitive area
Touching the product or parts thereof with bare
hands increases the desorption rate.
Always wear clean, lint-free gloves and use
clean tools when working in this area.

Vent the vacuum system.

Put the gauge out of operation.

Unfasten the lock screws and unplug the sensor cable.

(If you are using HPG400-SD or -SP, unfasten and un-
plug the interface connector, too (

→  [1] and [2]).

Remove the gauge from the vacuum system and install

the protective lid.



Maintenance, Repair

In case of severe contamination or a malfunction, the sensor
can be replaced (

→  [1]).

Gauge failures due to contamination or wear and

tear, as well as expendable parts (e. g. filament),
are not covered by the warranty.


Returning the Product

WARNING

WARNING: forwarding contaminated products
Contaminated products (e.g. radioactive, toxic,
caustic or biological hazard) can be detrimental
to health and environment.
Products returned to INFICON should preferably
be free of harmful substances. Adhere to the for-
warding regulations of all involved countries and
forwarding companies and enclose a duly com-
pleted declaration of contamination

1)

1)

Form under www.inficon.com

Products that are not clearly declared as "free of harmful
substances" are decontaminated at the expense of the
customer.
Products not accompanied by a duly completed declaration
of contamination are returned to the sender at his own
expense.


Disposal

DANGER

DANGER: contaminated parts
Contaminated parts can be detrimental to health
and environment.
Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant
regulations and take the necessary precautions
when handling contaminated parts.

WARNING

WARNING: substances detrimental to the en-
vironment
Products or parts thereof (mechanical and elec-
tric components, operating fluids etc.) can be de-
trimental to the environment.
Dispose of such substances in accordance with
the relevant local regulations.

Separating the components

After disassembling the product, separate its components
according to the following criteria:

• Contaminated components

Contaminated components (radioactive, toxic, caustic, or
biological hazard etc.) must be decontaminated in accord-
ance with the relevant national regulations, separated ac-
cording to their materials, and recycled.

• Other components

Such components must be separated according to their
materials and recycled.



Further Information

[1] www.inficon.com

Operating Manual
High Pressure / Pirani Gauge
HPG400, HPG400-SD, HPG400-SP
tina31d1 German
tina31e1 English
INFICON AG, LI–9496 Balzers, Liechtenstein

[2] www.inficon.com

Instruction Sheet
High Pressure / Pirani Gauge
HPG400-SD, HPG400-SP
tima32d1 German
tima32e1 English
INFICON AG, LI–9496 Balzers, Liechtenstein


EC Declaration of Conformity

We, INFICON, hereby declare that the equip-
ment mentioned below complies with the pro-
visions of the Directive relating to electromag-
netic compatibility 2004/108/EC.


High Pressure / Pirani Gauge

HPG400
HPG400-SD
HPG400-SP

Standards

Harmonized and international/national standards and specifi-
cations:
• EN 61000-6-2:2005

(EMC: generic immunity standard)

• EN 61000-6-3:2007

(EMC: generic emission standard)

• EN 61010-1:2001

(Safety requirements for electrical equip-

ment for measurement, control and laboratory use)

• EN 61326:2006

(EMC requirements for electrical equipment

for measurement, control and laboratory use)


Manufacturer / Signatures

INFICON AG, Alte Landstraße 6, LI-9496 Balzers

11 October 2011

11 October 2011

Dr. Urs Wälchli
Managing Director

Claudio Christoffel
Product Manager



LI–9496 Balzers
Liechtenstein
Tel +423

/

388

3111

Fax +423 / 388 3700
[email protected]

www.inficon.com

Advertising