INFICON HPG400-SP High Pressure / Pirani Gauge User Manual

High pressure / pirani gauge, Safety, Hpg400-sd devicenet

Advertising
background image

High Pressure / Pirani Gauge

With Fieldbus Interface

HPG400-SD
HPG400-SP

Instruction Sheet

tima32e1-a (2004-05)

About this document

This document is a supplement to the standard Instruction
Sheet enclosed with the HPG400 (

[1]). It contains spe-

cific information on using the HPG400-SD and HPG400-SP
gauge types in connection with a fieldbus interface and
should be used together with the standard Instruction
Sheet.

The symbol (

→ [XY]) refers to documents and files

listed under "Further Information".

Product Identification

In all communications with INFICON, please specify the in-
formation on the product nameplate. For convenient refer-
ence copy that information into the space provided below.

Model:
PN:
SN:
V W

INFICON AG, LI-9496 Balzers

Validity

This document applies to products with the following part
numbers:

HPG400-SD (DeviceNet):

353-527

(DN 25 ISO-KF)

353-528

(DN 40 CF-R)

HPG400-SP (Profibus):

353-525

(DN 25 ISO-KF)

353-526

(DN 40 CF-R)

The part number (PN) can be taken from the product name-
plate.

If not indicated otherwise in the legends, the illustrations in
this document apply to the vacuum connection
DN 25 ISO-KF. They apply to other vacuum connections by
analogy.

We reserve the right to make technical changes without prior
notice.

All dimensions in mm.

Trademarks

DeviceNet™ Open DeviceNet Vendor Association, Inc.

Intended Use

The High Pressure / Pirani Gauges of the type HPG400-SD
and -SP have been designed for vacuum measurement of
non-flammable gases and gas mixtures in a pressure range
of 2×10

-6

… 1 mbar. The control range of the gauge allows a

trend display from

<1 mbar (5×10

-2

mbar) ... 1000 mbar.

Safety

Symbols Used

DANGER

Information on preventing any kind of physical injury.

WARNING

Information on preventing extensive equipment and
environmental damage.

Caution

Information on correct handling or use. Disregard can lead
to malfunctions or minor equipment damage.

Personnel Qualifications

Skilled personnel

All work described in this document may only be carried out
by persons who have suitable technical training and the
necessary experience or who have been instructed by the
end-user of the product.

General Safety Instructions

All safety instructions given in

[1] and [2] apply to the

sensor types described in this document, too.

Liability and Warranty

INFICON assumes no liability and the warranty becomes null
and void if the end-user or third parties
• disregard the information in this document
• use the product in a non-conforming manner
• make any kind of interventions (modifications, alterations

etc.) on the product

• use the product with accessories not listed in the corre-

sponding product documentation.

The end-user assumes the responsibility in conjunction with
the process media used.

HPG400-SD DeviceNet

General Information

The HPG400-SD gauge has a fieldbus interface that
conforms to the DeviceNet standard (

→ [8]).

Via this interface, the following and further data are
exchanged in the standardized DeviceNet Protocol
(

→ [3], [8]):

• Pressure reading
• Pressure unit (mbar, Torr, Pa)
• Gauge adjustment
• Status and error messages

Two adjustable switching functions are integrated in the
gauge. The corresponding relay contacts are available at the
sensor cable connector.

The basic sensor and sensor electronics of the HPG400-SD
type are the same as in the standard HPG400 (

→ [1], [2]).

Technical Data HPG400-SD

General technical data of the sensor and sensor
electronics

→ [1], [2].

Fieldbus Interface

Fieldbus name

DeviceNet

Standard applied

→ [8]

Communication protocol,
data format

→ [3], [8]

Interface, physical

CAN bus

HPG400-SD DeviceNet parameters

Data rate
(adjustable via "RATE" switch)

125 kbaud
250 kbaud
500 kbaud
"P" (programmable
125 kbaud, 250 kbaud,
500 kbaud via DeviceNet)

(

→ [3])

Node address (MAC ID)
(adjustable via "ADDRESS,
MSD, LSD" switches)

0 … 63

dec

"P" (programmable
0 … 63 via DeviceNet)

(

→ [3])

DeviceNet connector

Micro-Style, 5 pins, male

Cable

Shielded special Device-
Net cable, 5 conductors
(

→ [8], [6])

Cable length, system wiring

According to DeviceNet
specifications (

→ [8],

[6])

Supply Voltages

DANGER

The gauge may only be connected to power
supplies, instruments or control devices that
conform to the requirements of a grounded pro-
tective extra-low voltage (SELV-E according to
EN 61010). The connection to the gauge has to
be fused.

Supply voltage at the sensor
connector, Pin 8

+24 VDC (+20 … 28 V)

Power consumption

<16 W

DeviceNet operation requires an additional, separate
power supply.

Supply voltage at the DeviceNet
connector, Pin 2

+24 VDC (+11 … 25 V)

Power consumption

<2 W

The gauge is protected from reversed polarity of the supply
voltage.

Advertising
This manual is related to the following products: