Sensor Cable Connection

For reasons of compatibility, the expression "sensor
cable" is used in this document, although the pres-
sure reading of the SD-type gauge is normally trans-
mitted via the DeviceNet interface.

Connector

D-Sub, 15 pins, male

Cable

Max. 15 conductors,
shielded

Cable length
(cross section/conductor)

≤35 m

(0.25 mm

2

)

≤50 m

(0.34 mm

2

)

≤100 m

(1.0 mm

2

)

Switching functions

2
Setpoints adjustable via
potentiometers (Setpoints
A and B), one floating,
normally open contact per
setpoint

Relay contact rating

≤60 VDC, ≤0.5 A

Gauge identification

56 k

Ω between Pin 10

(sensor cable) and GND

Grounding principle

→ "Electrical Connection"

Dimensions [mm]

Housing and vacuum connection

→ [1], [2]

Weight

353-527
353-528

430 g
695 g

Electrical Connection

Make sure the vacuum connection is properly made
(

→ [1], [2], " Vacuum Connection ").

n

If no sensor cable is available, make one according to

the diagram.

Threshold SP A

Threshold SP B

SP A

SP B

TxD

RxD

+U

b

Measuring signal

56 kΩ

3

6

1

4

9

11

13

14

8

2

12

10

5

15

1.25 AT

24V

Identification

0V

RS232

1

8

9

15

D-Sub, 15 pins, female
soldering side

Pin 1

Relay switching function A, COM contact

Pin 2

Signal output (measuring signal) 0 … +10 V

Pin 3

Threshold (Setpoint) A

0 … +10 V

Pin 4

Relay switching function A, N.O. contact

Pin 5

Supply common, GND

Pin 6

Threshold (Setpoint) B

0 … +10 V

Pin 7

Not connected

Pin 8

Supply voltage

+24 V

Pin 9

Relay switching function B, COM contact

Pin 10

Gauge identification

Pin 11

Relay switching function B, N.O. contact

Pin 12

Signal common, GND

Pin 13

RS232 TxD

Pin 14

RS232 RxD

Pin 15

Shielding, housing, GND

o

If no DeviceNet cable is available, make one according

to the following indications:

3

1

4

2

5

Micro-Style, 15 pins,
(DeviceNet), female
soldering side

Pin 1

Drain

Pin 2

Supply

+24 VDC

(DeviceNet interface only)

Pin 3

Supply common

GND

(DeviceNet interface only)

Pin 4

CAN_H

Pin 5

CAN_L

p

Connect the sensor cable to the gauge.

q

Connect the DeviceNet cable to the gauge.

Sensor cable

DeviceNet cable

r

Secure the sensor cable connector on the gauge using

the lock screws.

s

Lock the DeviceNet cable connector.

The gauge can now be put into operation.

Operation

Caution

Caution: data transmission errors

If the gauge is operated via an RS232 interface
(e.g. INFICON Vacuum Gauge Controller
VGC4XX) and the DeviceNet at the same time,
data transmission errors may occur.

The gauge must not be operated via an RS232
interface and the DeviceNet at the same time.

Operating Software

Before the gauge is put into operation, it has to be configured
for the DeviceNet. A configuration tool and the device specific
EDS file (Electronic Data Sheet) are required for this pur-
pose. This software can be downloaded via internet
(

→ [5]).

Node Address Setting

Set the node address (0 … 63

dec

) via the

"ADDRESS" "MSD" and "LSD" switches.
The node address is polled by the firmware
when the gauge is switched on. If the set-
ting deviates from the stored value, the new
value is taken over into the NVRAM. If a
setting higher than 63 is made, the previous
node address setting remains valid.

If the MSD switch is in the "P" position, the
node address is programmable via the
DeviceNet (

→ [3]).

Data Rate Setting

By means of the "RATE" switch, the data rate
can be set to 125 ("1"), 250 ("2") or 500 kbaud
("5").

If the switch is in any of the "P" positions, the
data rate is programmable via the DeviceNet
(

→ [3]).

The admissible data rate depends on several factors
(system parameters, cable lengths etc.

→ [8]).

Adjusting the Gauge

→ Adjustment and settings.

Setting the Switching Functions

→ Adjustment and settings.

The gauge is now ready for operation.

Status Lights

"STATUS MOD" (gauge status):

Light status

Meaning

Dark

No supply

Flashing
red/green

Selftest

Green

Normal operation

Red

Non recoverable error

"STATUS NET" (network status):

Light status

Meaning

Dark

Gauge not online:

– Selftest not yet concluded
– No supply,

→ "STATUS MOD" light

Flashing green

Gauge online but no connection:

– Selftest concluded, but no connection

to other nodes established

– Gauge not assigned to any master

Green

Gauge online; necessary connections
established

Flashing red

One or several input/output connections in
"timed out" status

Red

Communication error. The gauge has de-
tected an error that impedes communica-
tion via the network (e.g. two identical node
addresses (MAC ID) or "Bus-off")

t i ma 3 2 e 1 - a

(2004-05)

Original: German tima32d1-a (2004-5)