Adjustment and Settings

For HPG400-SD and HPG400-SP gauge

Adjusting the Gauge (Full Scale)

The gauge is factory calibrated. If used under different cli-
matic conditions, at extreme temperatures, through aging or
contamination and after exchanging the sensor, the charac-
teristic curve can be offset and readjustment can become
necessary. Only the Pirani element can be adjusted and only
at atmosphere.

Readjustment becomes necessary if
• at atmosphere the output voltage is <9.75 V or the display

reading is <atmosphere

• when venting the vacuum system, the output voltage

reaches 9.75 V before the measured pressure has
reached atmosphere.

(For more details

→ [2], "Maintenance, Repair").

n

If you are using a seal with centering ring and filter,

check that they are clean or replace them if necessary
(

→ [1], [2], "Deinstallation").

o

Put the gauge into operation.

Operate the gauge for

≈10 minutes at atmos-

pheric pressure. If the gauge was operated
within the hot cathode range, a cooling-down
time of

≈30 minutes is to be expected (gauge

temperature = environmental temperature).

p

Adjusting the gauge:

Insert a Pin (

≈ø1.3 mm) through the opening marked

"FULL SCALE" and push the button inside for at least
5 seconds.

Max. ø1.3

The gauge is now automatically adjusted. The
adjustment takes

≈4 s.

The gauge is now adjusted.

Setting the Switching Functions

The threshold values of switching functions A and B can be
set within the pressure range 2×10

-6

mbar … 100 mbar via

potentiometers "SETPOINT A" and "SETPOINT B".

For the corresponding threshold voltages, the following equa-
tion applies:

U

Threshold

= 10 / 9 × (log p

Setpoint

– c + 6)

Constant c depends on the pressure unit,

Where

p

c

[mbar]

0

[Torr]

-0.125

[Pa]

2

Pressure p

Me

asur

ed valu

e

Off

Off

On

Time t

Switching function

Hysteresis
10 % U

Threshold

Threshold

U

The hysteresis of the switching functions is 10% of the
threshold setting.

n

Put the gauge into operation.

o

Connect the (+) lead of a voltmeter to the threshold

measurement point of the selected switching function
("Setpoint A" Pin 3, "Setpoint B" Pin 6) and its (–) lead
to Pin 5.

Max. ø2.5

Setpoint A Pin 3
Setpoint B Pin 6

Pin 5

p

Using a screwdriver (max. ø2.5 mm), set the threshold

of the selected switching function (Setpoint A, B) to the
desired value U

Threshold

The setting of the switching functions is now concluded.
A functional check of the switching functions (On/Off) is
only possible via fieldbus interface (

→ [3] for

HPG400-SD,

→ [4] for HPG400-SP) or by measur-

ing the relay contacts with a continuity checker/ohm-
meter (

→ "Electrical Connection", sensor cable con-

nector).

Further Information

[1]

www.inficon.com
Instruction Sheet
High Pressure / Pirani Gauge HPG400
tima31e1
INFICON AG, LI–9496 Balzers, Liechtenstein

[2]

www.inficon.com
Operating Manual
High Pressure / Pirani Gauge
HPG400, HPG400-SD, HPG400-SP
tina31e1
INFICON AG, LI–9496 Balzers, Liechtenstein

[3]

www.inficon.com
Communication Protocol
DeviceNet™ High Pressure / Pirani Gauge
HPG400-SD
tira03e1
INFICON AG, LI–9496 Balzers, Liechtenstein

[4]

www.inficon.com
Communication Protocol
Profibus
Bayard-Alpert Pirani Gauge BPG400-SP
High Pressure / Pirani Gauge HPG400-SP
tira36e1
INFICON AG, LI–9496 Balzers, Liechtenstein

[5]

www.inficon.com
("Semiconductor and Vacuum coating processes,
Vacuum Gauges")
Product descriptions and downloads
INFICON AG, LI–9496 Balzers, Liechtenstein

[6]

www.odva.org
Open DeviceNet Vendor Association, Inc.
DeviceNet™ Specifications

[7]

www.profibus.com
Profibus user organization

[8]

European Standard for DeviceNet EN 50325

[9]

European Standard for Profibus EN 50170

LI–9496 Balzers
Liechtenstein
Tel +423 / 388 3111
Fax +423 / 388 3700
[email protected]

www.inficon.com

Advertising
This manual is related to the following products: