General information validity functional principle – INFICON Cube CDGsci (RS232C) User Manual

Page 2

Advertising
background image

2

tira90e1 (2014-01) RS232C.cp

The RS232C Serial Interface for Cube permits the communication between this
digital INFICON Capacitance Diaphragm Gauge and
• an appropriate controller or
• a

computer.


The RS232C Serial Interface integrated in the Capacitance Diaphragm Gauge
allows to digitally transmit measurement values and information on the gauge
status as well as to make parameter settings.

Caution

Caution: data transmission errors

Any attempt to simultaneously operate the gauge via the RS232C
Serial Interface and a fieldbus interface (DeviceNet or Profibus) or the
diagnostic port may result in incorrect data and data transmission
errors.

Therefore, it is inadmissible to simultaneously operate the gauge via
the RS232C Serial Interface and DeviceNet, Profibus, or the
diagnostic port.


This document applies to products with part numbers

3 C S 1 - 3 1 1 - 2 3 0 0

Flange

1
3

E

DN 16 ISO-KF
DN 16 CF-R
8VCR female

3
6
9

C

F

0.1
1
10
100
1000

Measurement range

(F.S.)

The part number (PN) can be taken from the product nameplate.

If not indicated otherwise in the legends, the illustrations in this document corres-
pond to CDGsci gauges with the DN 16 ISO-KF vacuum connection.

We reserve the right to make technical changes without prior notice.

The RS232C Serial Interface is used in duplex operation. The gauge continuously
(approximately every 100 ms) transmits a nine byte send string without request.

Instructions to the gauge are transmitted via five-byte receipt strings.

The RS232C Serial Interface is only available on the DIAG port.

• binary
• 8 data bits
• 1 stop bit
• no parity bit
• no

handshake