1 interface protocol, 1 send string – INFICON Cube CDGsci (RS232C) User Manual

Page 4

Advertising
background image

4

tira90e1 (2014-01) RS232C.cp

1 Interface

Protocol


The complete send string (frame) is nine bytes (byte 0 … 8) long. Bytes 1 … 7 form
the data string.

Byte No.

Function

Value

Comment

0

Data string length

7

Constant value

1

Page No.

4

Constant value for CDGsci

1)

2 Status

→ "Status byte"

3 Error

→ "Error byte"

4

Measured value high byte

<value

>

→ "Calculation of pressure value"

5

Measured value mid byte

<value

>

→ "Calculation of pressure value"

6 Read

command

<value

>

Read

value

7

Measured value low byte

1)

<value

>

→ "Calculation of pressure value"

8 Checksum

→ "Synchronization"

1)

The "Sensor Type" that is accessible via Byte 7 in all other CDGs can now be deduced
from address 56 (Exponent) and 57 (Mantissa).

Bit 0

Definition

0

Continuous output of measured value

1

Individual measured value (polling)

1)

1)

→  8, "DataTxMode". Send a read command of any parameter to the gauge for
requesting a send string.

Bit 2

Bit 1

Definition

1

0

Manual setpoint setting

1

1

Zero adjust active

Bit 3

Definition

0

⇔ 1

Toggle bit, changes with every string
received correctly

Bit 5

Bit 4

Definition

0

0

Current pressure unit mbar

0

1

Current pressure unit Torr

1

0

Current pressure unit Pa

Bit 6

Definition

0

Standard measurement mode

1

Reserved for internal use

Bit 7

Definition

0

Heating

1

Sensor temperature attained


1.1 Send String

Structure of send string

Status byte
(byte No. 2)

Advertising