12

tira90e1 (2014-01) RS232C.cp



Parameter name

Data type

Byte No. 1 Byte No. 2

Byte No. 3

Comment

Reset

uint8 / W

0x40

0x00

0x00

Power reset: Starts continuous pressure output

ResetFactory

uint8 / W

0x40

0x01

0x00

Factory reset: Sets factory configuration

ZeroAdjust

uint8 / W

0x40

0x02

Starts zero offset adjustment

W = Write


<data_value>×a

Setpoint_level xy =

b

× F.S.R._Mantisse×10

(F.S.R._Exp)


Parameter

1)

Description

Setpoint_level xy Setpoint threshold in the selected pressure unit.

1)

Further

parameter

→  13, table "Parameter"

Minimum value lower switching threshold = 0 ; negative values are not

admissible.

Maximum value lower switching threshold = F.S.R. – 1 % hysteresis.

Zero_Adjust_Value contains the zero pressure offset value required for zeroing
(writable and readable).

• Automatic zero_adjust function via key or command (→ table "Variables for

bytes No. 2 and 3 (special services)").

• Base-Pressure-Adjust for adjusting a defined zero offset, e.g. if the required

final pressure as indicated in the operating manual is not reached.

The Zero_Adjust_Value consists of the high and low byte and has to be converted
with the "Pressure value" formula (

→  5).

<data_value>×a

Zero_Adjust_Value =

b

×F.S.R._Mantisse×10

(F.S.R._Exp)


Parameter

1)

Description

Zero_Adjust_Value Zero pressure offset in the selected pressure unit

(

→  13, table "Parameter").

1)

Further

parameter

→  13, table "Parameter"


Maximal remaining offset value. The Zero_Adjust can only be executed within this
value.

Variables for bytes No. 2 and 3
(special services)

Description of variables

Setpoint_level xy

Zero_Adjust_Value

Remaining_Zero