JULABO EH-39 Open Bath Circulator User Manual

Page 19

Advertising
background image

EH

19


Warning:

Tubing:

At high working temperatures the tubing used for temperature application and
cooling water supply represents a danger source.
A damaged tubing line may cause hot bath fluid to be pumped out within a short
time.

This may result in:

 Burning of skin

 Difficulties in breathing due to hot atmosphere

Safety recommendations

 Employ suitable connecting tubing.
 Make sure that the tubing is securely attached.
 Avoid sharp bends in the tubing, and maintain a sufficient distance from

surrounding walls.

 Regularly check the tubing for material defects (e.g. for cracks).
 Preventive maintenance: Replace the tubing from time to time.


5.5.

Adjusting the pump flow

1

2

The pump flow is pre-adjusted in the factory and can be
modified to suit user requirements.

 Using a screwdriver turn the screw (1) anti-clockwise by 360

°.

 Using flat pliers turn the marking of the slide (2) to the

desired position.

 Tighten the screw.

Examples:

extern 80 %

intern 100 %

intern 20 %

A

B

D

C

Internal applications in the bath

A 100 % internal bath circulation (for large bath tanks)
B Reduced internal bath circulation

(for smooth surface of bath fluid)

External/internal applications

C 40 % external discharge,

60 % internal circulation
(for large bath tanks)

D 80 % external discharge,

20 % internal circulation
(for small bath tanks)