1. Intended

use

JULABO circulators have been designed to control the temperature of specific fluids in a bath
tank. The units feature pump connections for temperature control of external systems (loop
circuit).

JULABO circulators are not suitable for direct temperature control of foods,
semi-luxury foods and tobacco, or pharmaceutical and medical products.
Direct temperature control means unprotected contact of the object with the
bath medium (bath fluid).

1.1. Description

PID1

°C

40

80

160

120

0

The circulators are operated via the splash-proof keypad. The implemented
microprocessor technology allows to set and to store the setpoint that can
be indicated on the LED temperature display.

The PID temperature control adapts the heat supplied to the thermal
requirements of the bath.

ATC - Absolute Temperature Calibration (1-point calibration)

Safety installations conforming to IEC 61010-2-010
The excess temperature protection is a safety installation independent from
the control circuit.
The safety value is set using a tool (screwdriver).
If the low level protection device is triggered, a complete shutdown of the
heater and circulating pump is effected.

Electrical connection:
Alarm output for external alarm message or control of JULABO refrigerating
baths.

2. Operator responsibility – Safety recommendations

The products of JULABO ensure safe operation when installed, operated, and maintained
according to common safety regulations. This section explains the potential dangers that may
arise when operating the circulator and also specifies the most important safety precautions to
preclude these dangers as far as possible.
 The operator is responsible for the qualification of the personnel operating the units.
 The personnel operating the units should be regularly instructed about the dangers involved

with their job activities as well as measures to avert these dangers.

 Make sure all persons tasked with operating, installing, and maintaining the unit have read

and understand the safety information and operating instructions.

 When using hazardous materials or materials that could become hazardous, the circulator

may be operated only by persons who are absolutely familiar with these materials and the
circulator. These persons must be fully aware of possible risks.