Operating procedures

22

6.3.

Automatic / non-automatic start mode

Keep depressed the

key and

turn on the circulator with the mains power switch.

For a short while the LED temperature DISPLAY indicates the
effective start mode:

AUTOSTART on.

AUTOSTART off.

NOTE:
The circulator has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start
mode, that the unit must enter a safe operating state after a power
failure (non-automatic start mode). This safe operating state is
indicated by „OFF“ on the LED temperature display. A complete
shutdown of the main functional elements such as heater and
circulating pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the
start of the circulator directly by pressing the mains power switch or
using a timer.

Warning:

For supervised or unsupervised operation with the AUTOSTART function, avoid
any hazardous situation to persons or property.
The circulator does no longer conform to N.A.M.U.R. recommendations.

Please make sure that all safety devices of the unit have been set properly.

6.4.

Setting the temperatures

Factory setting:

25°C

Setting can be carried out in the start/stop condition.

1. Press one of the keys

for a short moment. The

setpoint value instead of the actual value is indicated on the
display for about 8 seconds. The value can now be changed.

2. Change value:

Press

to set a higher value.

Press

to set a lower value.

Keep the keys depressed for the value to change fast.

3. Press the

key to store the value.