JULABO F26-ED Refrigerated / Heating Circulator User Manual

Page 10

Advertising
background image

Operator responsibility – Safety recommendations

10

Safety installations according to IEC 61010-2-010:

Excess temperature protection

adjustable from 20 to 120 °C

Low liquid level protection

float switch

Classification according to DIN 12876-1

class I

Alarm indication

optical + audible (permanent)

Environmental conditions according to EN 61 010, part 1:

Use only indoor.

Altitude up to 2000 m - normal zero.

Ambient temperature: +5 ... +40 °C (for storage and transportation)

Air

humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,

linear decrease down to 50 % relative humidity at a temperature of +40 °C

Max. mains fluctuation of ±10 % are permissible.


Protection class according to EN 60 529

IP21

The unit corresponds to Class I

Overvoltage category II

Pollution degree 2

Caution:

The unit is not for use in explosive atmosphere


Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).

Advertising
This manual is related to the following products: