2.4. Technical

specifications

F12-ED

F25-ED

Working temperature range

°C

-20 ... 100

-28 ... 100

Temperature stability

°C

±0,03

±0,03

Temperature selection

digital

digital

Temperature indication

LED

LED

Resolution

°C

0.1

0.1

Temperature control

PID1

Heater wattage (at 230 V)

kW

2,0

2,0

Heater wattage (at 115V)

kW

1,0

1,0

Cooling capacity

Medium ethanol

°C

kW

+20 0 -20

0.16 0.1 0.02

+20 0 -20

0.26 0.2 0.06

Refrigerant

R134a

R134a

Circulating pump:

discharge, max.at 0 bar

l/min

15

15

pressure, max. at 0 l

bar

0.35

0.35

Overall dimensions (WxDxH)

cm

20x36x56

23x42x61

Bath opening (WxL)

cm

13x15

12x14

Bath depth

cm

13

14

Filling volume

liters

3 ... 4,5

3 ... 4,5

Weight

kg

22

30

Ambient temperature

°C

5 ... 40

5 ... 40

Mains power connection 230 V/50 Hz

V/ Hz

207-253 / 50

207-253 / 50

Current input (bei 230 V)

A

11 (CH 9 + 1)

12 (CH 9 + 2)

Mains power connection 230 V/60 Hz

V/ Hz

207-253 / 60

207-253 / 60

Current input (at 230 V)

A

11

12

Mains power connection 115 V/60 Hz

V/ Hz

103-127 / 60

103-127 / 60

Current input (at 115 V)

A

12

13

Mains power connection 100V/50-60Hz

V/ Hz

90-110 / 50-60

90-110 / 50-60

Current input (at 100 V)

A

15

13

All measurements have been carried out at:

rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.