JULABO F26-ED Refrigerated / Heating Circulator User Manual

Page 12

Advertising
background image

Safety notes for the user

12

 Do not stay in the area below the unit.
 Make sure you read and understand all instructions and safety precautions

listed in this manual before installing or operating your unit.

 Never operate the unit without bath fluid in the bath.
 Check the filling level of the bath fluid from time to time. Pump and heater

must always be fully covered with the bath fluid!

 The instrument is not suited for unsupervised continuous operation.
 Do not drain the bath fluid while it is hot!

Check the temperature of the bath fluid prior to draining (by switching the
unit on for a short moment for example).

 Observe the limited working temperature range when using plastic bath

tanks.

 Employ suitable connecting tubing.

Make sure that the tubes are securely attached.

 Never operate damaged or leaking equipment.
 Always turn off the unit and disconnect the mains cable from the power

source before performing any service or maintenance procedures, or
before moving the unit.

 Always empty the bath before moving the unit.
 Never operate equipment with damaged mains power cables.

 Some parts of the bath cover and the pump connections may become

extremely warm during continuous operation. Therefore, exercise particular
caution when touching these parts.


Notice:

Check the safety installations at least twice a year!

 Excess

temperature

protection

according to IEC 61010-2-010.

With a screwdriver turn back the adjustable excess temperature protection
until the shut-down point (actual temperature).

 Low level protection according to IEC 61010-2-010.

To check the function of the float, it can be manually lowered with a
screwdriver for example.