21

6. Operating

procedures

6.1. Power

connection

Caution:

 Only connect the unit to a power socket with earthing contact (PE –

protective earth)!

 The power supply plug serves as safe disconnecting device from the line

and must be always easily accessible.

 Never operate equipment with damaged mains power cables.

 Regularly check the mains power cables for material defects (e.g. for

cracks).

 We disclaim all liability for damage caused by incorrect line voltages!



Check to make sure that the line voltage matches the supply
voltage specified on the identification plate.

 Connect the circulator with mains power cable (10a) to the

mains outlet (17).

 Connect the refrigerated circulator with mains power cable

(10b) to the mains socket.

6.2. Switching

on / Start – Stop

Switching on:

Circulator and cooling machine may be turned on and off with
separate mains switches. The integrated control light will
illuminate to indicate that power has been applied.

 The unit performs a self-test. The segments of the 4-digit LED

temperature DISPLAY and all indicator lights will illuminate

(as

illustrated on the left).

Then the software version (example: 11.0) appears. The
display „OFF“ indicates the unit is ready to operate. (standby
mode).

Start:

 Press the

key for about 4 seconds.

The LED temperature DISPLAY indicates the actual bath
temperature.

Stop:
 Press

the

key for about 4 seconds.

 Turn the unit off with the mains power switch.