10


Recirculating Cooler
Cooling

FL2506 FLW2506
air cooled

water cooled

Working temperature range

°C

-20 ... +40

-20 ... +40

Temperature stability

°C

±0.5

Temperature selection:

digital

via key pad

indication

on

LED-DISPLAY

remote control via personal computer

indication on monitor

Temperature indication:

LED-DISPLAY

Resolution

°C

0.1

Temperature control

PID

1

Temperature sensor

Pt 100

Excess temperature protection

85 °C - fixed value

Low liquid level protection

float switch

Cooling capacity
Medium: Mixture water-glycol

°C
kW

+20 0 -10

+20 0 -10

2.5 1.0 0.3

2.5 1.0 0.3

Cooling compressor

1- stage

Refrigerant

R404A

Electrical connections:

Computer

interface

RS232

Alarm

output

for external alarm signal

Circulating pump:

discharge, max.

at 0 bar

l/min

60

60

pressure, adjustable at 0 Liter

bar

0.5 ... 6.0

0.5 ... 6.0

Feed pressure indication

bar

Manometer

Manometer

Filling level indicator

sight

glass

Filling volume

from ... to

liters

24 … 30

Dimensions (WxLxH)

cm

60x76x115

Weight

kg

158

160

Ambient temperature range

°C

5 ... 40

Return flow temperature

°C

80

max.

Cooling water

Flow rate at 20 °C inlet temperature


l/min


-----

4.1

IP class according to IEC 60 529

IP

21

Mains power connection 230 V/50 Hz

V/ Hz

207-253 / 50

207-253 / 50

Current draw at 230 V

A

14

15

Mains power connection 208 - 220 V/60 Hz

V/ Hz

197 - 242 / 60

197 - 242 / 60

Current draw at 208V / 220 V

A

16

------


All measurements have been carried out at: rated voltage and frequency,
ambient temperature: 20 °C