The out commands are valid only in remote control mode.


Examples:
Command to set the working temperature to15,5 °C:
out_sp_00

15.5

Command to ask for the working temperature
in_sp_00

Response from the recirculating cooler:
15.5

LF





11.3. List of commands

out commands: Setting parameters or temperature values.

Command

Parameter

Response of recirculating cooler

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_sp_00 xxx.xx

Set working temperature


in commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of recirculating cooler

version

none

Number of software version (V X.xx)

status none

Status message, error message (see page 30 )

in_pv_00

none

Actual bath temperature.

in_sp_00 none

Working

temperature

in_mode_05 none

Recirculating cooler in Stop/Start condition:
0 = Stop
1 = Start

11.4. Status

messages

Status messages

Description

00 MANUAL STOP

Recirculating cooler in „OFF“ state.

01 MANUAL START

Recirculating cooler in keypad control mode.

02 REMOTE STOP

Recirculating cooler in „r OFF“ state.

03 REMOTE START

Recirculating cooler in remote control mode.