Warning without a complete shutdown of the unit

Excess temperature warning starting at 75 °C
The return temperature soon reaches the swith-off value of the high
temperature protection (85 °C).

Cooling of the condenser is affected. (see page 32)

Clean air-cooled condenser.

Check the flow rate and cooling water temperature on water-cooled
condenser.

If the unit cannot be returned to operation, contact an authorized JULABO
service station.


Disturbances that are not indicated.

Overload protection:: a) for cooling machine

b) for pump motor

After a short cooling interval, the unit will automatically start running.

F1, F2, F3, F4

Mains circuit breakers (resettable) FL250x

2 Mains circuit breakers (resettable) 4 A

(for Electronics)

2 Mains circuit breakers (resettable) 10 A (for Pump motor)

0

1

F1, F2, F3, F4

Mains circuit breakers (resettable) FL400x

1 Motor protection switch for compressor motor

2 Mains circuit breakers (resettable) 4 A

(for Electronics)

2 Mains circuit breakers (resettable) 10 A (for Pump motor)