JULABO SW23 Shaking Water Baths User Manual

Page 22

Advertising
background image

Switching on

22

7.1. Setting the temperature











Display and adjustment of the working temperature:

Press the setpoint key

.

The indicator light blinks and the value previously set appears
on the MULTI-DISPLAY (LED).
(example: 25.0 °C).

If no further key is pressed the display will return to show the
actual bath temperature after approx. 8 seconds.

Use the cursor keys

to move left or right on the

display until the numeral you wish to change is blinking.

Use the increase/decrease arrows

to change the

selected numeral (0, 1, 2, 3, ... 9).

Press enter

to store the selected value

(example: 38.0 °C).

The working temperature is maintained constant after a short heat-
up time (e. g. 38.0 °C).

Notice:

When the working temperature is higher than 50 °C, it might happen that due to
strong production of steam there is considerable dripping on the inside of the lift-

up Makrolon

cover. Some drops may fall directly into the material to be

tempered.

 Always properly seal all sample containers.

7.2. Warning functions or temperature limit




......

More protection for your samples in the bath!
As soon as the actual temperature leaves one of the pre-adjusted
limits, this status is evaluated.

The high- and low-temperature limit can be evaluated in two ways
(see page 31).

1. As pure warning function with an acoustic signal in regular

intervals. (Signal - Pause)

2. As temperature limit by switching-off the heating and alarm.