12.2. List of commands

Command

Parameter

Response of the shaking water bath

version

none

Number of software version(V X.xx)

status

none

Status message, error message (see below)

out_mode_05

0

STOP - returns the water bath to the „OFF“ state

out_mode_05

1

START - water bath is switched to the operating state

out_mode_08

0

Circulator pump „OFF“

out_mode_08

1

Circulator pump „ON“

out_mode_09

0

Shaking operation drive motor „OFF“

out_mode_09

1

Shaking operation drive motor „ON“

out_sp_00

xxx.x

Set working temperature„T“

out_sp_02 xxx.x

Set high temperature warning limit

out_sp_03 xxx.x

Set low temperature warning limit

out_sp_16

xxx.x

Set shaking frequency

in_sp_00

none

Ask for working temperature „T“

in_sp_02 none

Ask for high temperature warning limit

in_sp_03 none

Ask for low temperature warning limit

in_sp_16

kein

Ask for shaking frequency

in_pv_00

none

Ask for actual bath temperature

in_pv_01

none

Ask for the heater wattage being used