38

12.3. Status messages

Message Description

01 MANUAL START

Waterbath in keypad control mode.

02 REMOTE STOP

Waterbath in „OFF“ state

03 REMOTE START

Waterbath in remote control mode

12.4. Error messages

Message Description

-01 TEMP / LEVEL ALARM

Safety temperature or low liquid level alarm

-03 EXCESS TEMPERATURE

WARNING

High temperature warning „

„.

-04 LOW TEMPERATURE WARNING

Low temperature warning „

„.

-05 TEMPERATURE MEASUREMENT

ALARM

Error in measuring system

-07 I2C-BUS WRITE ERROR

-07 I2C-BUS READ ERROR

-07 I2C-BUS READ/WRITE ERROR


Internal error

-08 INVALID COMMAND

Invalid command

-10 VALUE TOO SMALL

Entered value too small

-11 VALUE TOO LARGE

Entered value too large

-12 WARNING : VALUE EXCEEDS

TEMPERATURE LIMITS

Value lies outside the adjusted range for the high
and low temperature warning limits. But value is
stored.

-13 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode