JUMO 902210 Mineral-Insulated RTD Temperature Probes with Bare Connection Wires According to DIN EN 60751 Data Sheet User Manual

Technical data

Advertising
background image

2013-01-21/00073472

Data Sheet 902210

Page 1/10

JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14

36039 Fulda, Germany

Postal address:

36035 Fulda, Germany

Phone:

+49 661 6003-0

Fax:

+49 661 6003-607

E-mail:

[email protected]

Internet:

www.jumo.net

JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:

+44 1279 635262

E-mail:

[email protected]

Internet: www.jumo.co.uk

JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866

1-800-554-5866

Fax:

315-437-5860

E-mail:

[email protected]

Internet: www.jumousa.com

Mineral-Insulated RTD Temperature

Probe According to DIN EN 60751

• For temperatures between -50 (-200) to +600 °
• Flexible sheath cable with vibration-resistant measuring insert
• As single or double RTD temperature probe

in 2-wire, 3-wire, or 4-wire circuit

• Quick response time
• Application-specific insertion length

Due to their specific features, mineral-insulated RTD temperature probes are used in chemical
plants, power plants, pipelines, engine building, test rigs, and any measuring locations requiring
flexibility and simple replacement. Inside the flexible and thin-walled sheathed cable out of
stainless steel, the low ohmic conductor copper wires are embedded in compressed, heat-re-
sistant magnesium oxide.

The temperature sensor with 2-wire, 3-wire, or 4-wire technology is connected to the conductor
copper wires and installed in the protection tube out of stainless steel. The protection tube and
sheathed cable are welded to one another. The diameters already start at 1.9 mm.

The good thermal transfer between the protection tube and temperature sensor enables short
response times (t

0.5

from 0.7 s) and excellent measuring accuracy. The vibration-resistant con-

struction guarantees a long operating life. The flexible probe tube allows temperature measure-
ments to be made at measuring points that are difficult to access. The smallest bending radius
is 5 times the outer diameter.

The measuring insert is normally fitted with a Pt100 temperature sensor according to DIN EN
60751, class B in 2-wire circuit; versions with Pt500 or Pt1000 can be ordered as well. The con-
nection is also available in 3-wire and 4-wire circuits.

Technical data

Terminal head

Form B, diecast aluminum, M20 × 1.5; IP65, ambient temperature -40 to +100 °C
Form BUZ, diecast aluminum, M20 × 1.5; IP65, ambient temperature -40 to +100 °C
Form J, diecast aluminum, M16 × 1.5, IP65, ambient temperature -40 to +100 °C
Caution: reduced ambient temperature when using transmitter

Connection

Cable ends stripped bare; alternatively available with ferrules, plug-in sleeves, or multi-pole electrical
connectors

Connecting cable

Silicone, ambient temperature -50 to +180 °C
PTFE, ambient temperature -190 to +260 °C
Metal braiding, ambient temperature -50 to +350 °C

Process connection

Thread, stainless steel 1.4571

Protection tube

Stainless steel 1.4571, Ø 1.9 mm, Ø 3 mm, and Ø 6 mm

Adapter sleeve (end closure)

As standard, the adapter sleeve (end closure) is designed up to 120 °C; it is available up to 300 °C
upon request.

Measuring insert

Pt100/Pt1000 temperature sensor, DIN EN 60751, class B in 2-wire, 3-wire, or 4-wire circuit

Response times

In water with 0.4 m/s / in air with 3 m/s
Ø 1.9 mm: water t

0.5

= 0.7 s, t

0.9

= 2.1 s/air t

0.5

= 7.2 s, t

0.9

= 20.5 s

Ø 3.0 mm: water t

0.5

= 1.3 s, t

0.9

= 4.0 s/air t

0.5

= 13.5 s, t

0.9

= 41.0 s

Ø 6.0 mm: water t

0.5

= 5.0 s, t

0.9

= 11.5 s/air t

0.5

= 37.5 s, t

0.9

= 117.5 s

Transmitter

Analog transmitter, 4 to 20 mA output, data sheet 707030
Programmable transmitter, 4 to 20 mA / 20 to 4 mA output, data sheet 707010
Programmable transmitter, 4 to 20 mA / 20 to 4 mA and HART

®

-interface output, data sheet 707010

Advertising