CHAPTER 4 TROUBLE SHOOTING


1 Failure Analysis and Processing Procedures

No.

Phenomenon

Failure Analysis

Troubleshooting

Disconnected power supply

Check power supply for correct
connection

Bad fuse

Replace it (250V 2.5A,Φ5×20mm
or 125V 5A, Φ5×20mm )

Switch failure

Replace switch

1

No display and the two beeps
missed when power on

Others

Contact Labnet or your distributor

Ventilation blocked

Clear the airflow blockage

Bad connection

Open the device, check the
connectors for reliable connection

2

Abrupt change in heating
speed or bad temperature
control accuracy

Cooling module failure

Contact Labnet or your distributor

Cooling module failure

3

Significant change in cooling
speed, or unavailable
temperature below ambient

Fan failed or stopped

Contact Labnet or your distributor

Bad sensor

4

Heating and cooling
terminated

Malfunction of all cooling
modules

Contact Labnet or your distributor

Bad contact between connectors

Loose connection between
connectors

Lid heater failure

5

Lid cannot be heated

Lid sensor failure

Contact Labnet or your distributor

Bad contact of chip with its
socket

6

Abnormal characters
displayed

Chip malfunctioning

Contact Labnet or your distributor

7

Inactive keys

Membrane panel damaged

Contact Labnet or your distributor