2 Error Messages

No.

Error Message

Possible Reasons

1

Error Message when running “Please contact the
distributor
”, with “AD1213 failure”

The microchip can not communicate with the
AD1213. AD1213 broken or circuit board trouble.

2

Error Message when running “Please contact the
distributor.
”, with “Heatsink temperature is
too high”

Heat

sink

temp.

higher

than

70℃,the

temperature sensor is broken, there is a short
circuit or the fan is not working.

3

Error Message when running “Please contact the
distributor.”
, and "Heatsink sensor is
disconnected."

Heat sink temp. less than 5℃,the sensor broken,
there is a short circuit or the fan is not working.

4

Error Message when running "Please contact the
distributor."
,and "Temperature sensor No.1 is
disconnected."

Block sensor No.1 reports an open circuit

5

Error Message when running "Please contact the
distributor."
,and "Temperature sensor No.1 is
short-circuited."

Block sensor No.1 reports a short circuit

6

Error Message when running "Please contact the
distributor."
,and

"Hotlid

sensor

is

disconnected"

Hot-lid sensor reports an open circuit

7

Error Message when running "Please contact the
distributor."
,and "Hotlid sensor is short-
circuited."

Hot-lid sensor reports a short circuit