Appendix 2 parameter values – Labnet MultiGene Mini Personal Thermal Cycler User Manual

Page 20

Advertising
background image

18

APPENDIX 2 Parameter Values

Item

Parameter

Scale

Range

Note

1

Temp(ºC)

Temp. Control point

4.0~99.0

--:--

set the value before

“:”through “Function”
key.)

The time of Temp. control is
infinite, meanwhile, the hotlid
turned off.

2

Time

Time of Temp.
Control(mm:ss)

00:00~59:59

0.1 ~5.0

3

Ramp(C/S)

Ramping Rate

ºC/Second)

#.#

Maximum of Ramping Rate

4

+Temp

Temp. Gain of Every
Cycler

0.0~9.9

5

+Time

Time Gain of Every
Cycler(mm:ss)

0:00~9:59

Cycle1

Section Numbers

1~5

×00

Cycler Numbers from
beginning to the end

1~99

From 00

The start segment of

the section

X(and existed segment

no.)

6

to 00

The end segment of

the section.

≤ X+15

Cycle1:×00 From 00 to 00

7

Grad.

Gradient Temp.
Difference( )

0 ~30

8

Date

Date(YY-MM-DD)

00(~99)-01(~12)-
01(~31)

year-month-date

9

Time

Time(HH:MM:SS)

00:00:00~23:59:59

hour:minute:second

10

Default File

The default file name

Max. length is 11
characters.

11

Default User

The default user
name

Max. length is 11
characters.

Block

Under

this

mode

(once

reaching target temperature
namely entering into the
thermostable state), the time
for

reagent’s

temperature

under target temperature is
less than setting time (see Fig.
1).

12

Control Mode

Control Mode

Tube

Under

this

mode

(once

entering into the thermostable
state after reaching target
temperature and decreasing
about

3°C),

reagent’s

temperature can be close to
target temperature very fast
(see Fig. 2).

13

Sample Volume

Sample Volume(ul) 10~200

State of Hotlid: OFF

OFF

14

Hotlid Temp

State of Hotlid:
Open( )

30~99

15

Block Temp

State of Block Temp
during the heating of
Hotlid(ºC)

0~99

16

Key Beep

Key Sound (ON or
OFF)

OFF/ON

Humming

17

Run End Beep

Run End Sound (ON
or OFF)

OFF/ON

Humming continuously

Advertising