Weekly cleaning – Leica Biosystems IPS User Manual

Page 15

Advertising
background image

15

Leica Printer IP C/S

Key Instructions V 2.0 - 08/2013

Cleaning the print head

The print head must be cleaned

manually once a week or if the

message "15" is displayed.

• Open the printer lid and then

press CLEAN and LOADED

simultaneously.

• The print head (72,

Fig. 7

)

moves upward from the re-

placement plate.

• Push the lever (69,

Fig. 8

) upwards. Then

remove the red replacement plate (70,

Fig. 8

).

• Take one of the provided Leica foam

swabs, wet it with some alcohol and

clean the print head. Be sure not to use

too much; alcohol must not drip into the

instrument.

+

Weekly cleaning

72

70

69

Fig. 8

Fig. 7

Fig. 9

Never use acetone or xylene!

Only 95 % or 100 % alcohol is per-

mitted for cleaning purposes.

• Clean the red replacement plate using a

Leica foam swab moistened with alcohol.

• The sealing lip of the replacement plate

(71,

Fig. 9

) has to be absolutely clean – NO

ink residue is allowed on the front and rear

side! Check the sealing lip for damage.

71

Advertising
This manual is related to the following products:

IPC