16

Leica Printer IP C/S

Key Instructions V 2.0 - 08/2013

If no button is pressed to ac-

knowledge the end of the clean-

ing procedure, the print head will

be closed automatically after a

few minutes to prevent it from

drying out. The message "15"

remains.

Never rotate the swab – this can

cause premature scratches and

wear on the nozzle plate of the

print head.

• Carefully insert the swab under the print

head (

Fig. 10

). Apply light pressure up-

wards on print head from the back right

to the front left (with respect to the sealing

lip of the replacement plate). Use a clean

surface of the swab each time you touch

it to the print head.

Fig.10

The printer automatically closes the print head after 2.5 minutes. A signal sounds

30 s before it automatically closes and a countdown appears on the display. To

prevent damage to the print head, wait 30 seconds and then press CLEAN and

LOADED again. End the cleaning process and press any button on the control

panel to conclude the cleaning.