Leica Biosystems IPS User Manual

Page 17

Advertising
background image

17

Leica Printer IP C/S

Key Instructions V 2.0 - 08/2013

Automated unload station

Fig. 12

• Remove the trays; clean glass dust,

splinters and debris (IP S) from the guides

and ejector.

• The trays themselves can be cleaned

with a household cleaner. Do not use any

solvents to clean the trays!

• Prior to reinserting them into the instru-

ment, the trays must be completely dry.

Loading station

Fig. 11

• Clean dust and debris from the ejection

unit of the magazines, magazine holders

and chute.

Ball bearing (only IP S)

Fig. 11

• Chute (2,

Fig. 11

)

• Transport station (1,

Fig. 13

)

• Drying module (3,

Fig. 15

)

Cleaning the instrument
• Observe the general cleaning instructions in Chapter 6 of the provided Instructions

for Use.

• Switch off the instrument before cleaning it!
• Always remove dust and debris using a brush or small vacuum cleaner!

Outer surfaces
• The painted outer surfaces (including those of the automated unload station) can be

cleaned with a mild cleaning agent and then wiped dry with a slightly moistened cloth.

Do NOT use solvent on the outer surfaces or the lid!

• Regularly remove dust on the ventilation slots on the rear side of the instrument.

Never use compressed air to clean – risk of injury!

1

Fig. 12

Fig. 11

1

2

Advertising