Dioptric correction and parfocality – Leica Biosystems M80 User Manual

Page 40

Advertising
background image

Leica M-series

User Manual

40

Adjusting

1. Set the dioptric correction for both

eyepieces to "0".

2. Select the lowest magnification and focus

on a flat specimen.

3. Select the highest magnification and read-

just the sharpness.

4. Select the lowest magnification again, but

do not look into the eyepieces.

6. Rotate the eyepieces counterclockwise

in the "+" direction as far as they will go
(+5 diopter settings).

Leica stereomicroscopes are parfocally matched.
The prerequisite for this is the correct setting of
the diopters and the parfocality. The following
adjustments only have to be carried out once
by each user.

Preparation

Move the lever of the video/phototube to
the "observation" position and open the
diaphragm.

7. Look into the eyepieces.

8. Slowly rotate each eyepiece individually

in the "–" direction until each eye sees the
object sharply imaged.

9. Select the highest magnification and refo-

cus if necessary.

Now, if you adjust the magnification from the
lowest to the highest level, the specimen is
always brought into sharp focus. If not, repeat
the process.

Dioptric Correction and Parfocality

Advertising
This manual is related to the following products:

M60, M50