Leica M-series

User Manual

7

Safety Regulations (cont'd.)

Light sources: safety regulations

Protective measures of the manufacturer:

★

UV protection screen in front of the speci-
men plane prevents the user from looking
directly into the UV rays.

★

Dummy filter carriers in the free positions
of the rapid filter changer prevent direct UV
radiation from reaching the eyes.

★

UV filters are installed in the observation
beam paths to protect the eyes.

★

The stray-light protection on the lamp
housing prevents irradiation of the hands.

Warning

★

UV radiation could damage the eyes.
Therefore:

★

Never look into the light spot on the speci-
men plane without a UV protection screen.

★

Never look into the eyepieces if no excita-
tion filter is in the beam path.

★

Fill empty filter positions with dummy filter
carriers (e.g. M165 FC).

★

Do not select a white, strongly reflective
background for the specimen.

Supply unit

Always unplug the supply unit from the power
supply:

★

When installing and disassembling the
lamp housing

★

Before opening the lamp housing

★

When replacing the high-pressure mercury
lamp and other parts, such as the heat-
absorbing filter or the collector

★

During maintenance work on the supply
unit