ALIBRATING THE

DI-100

Please ensure the UNIT and LC (load capacity) commands are entered accordingly, prior to
performing the calibration

.

4.1 Calibration

Mode

The DI-100 supports two calibration modes: The mV/V calibration or the 2-Point Linear Calibration.

You may perform the calibration for either mode independently and select the actual calibration
mode in use.

4.2

Millivolt-per-Volt (mV/V) Calibration

Obtain the Rated Output (in millivolt-per-volt, mV/V) from your load cell’s calibration certificate or
from the label attached on the load cell. This number is the best fit mV/V value determined during
the factory calibration procedure.

• To select the mV/V mode, type: CAL m <Enter>
• To enter mVolt calibration, type: MVOLT x.yyyy <Enter>

Where: x.yyyy is the mV/V rated output. The parameters are saved permanently on the DI-100
until you perform another millivolt calibration. You may start to use your load cell now:

• Type:

? <Enter>

to view the command list.

• To tare (i.e., reset the zero) the load cell, type: TARE <Enter>
• To view weight once, type: W <Enter>
• To view weight continuously, type: WC <Enter>
• Pull or push on the load cell, the weights should change accordingly.
• Press

<Enter>

to stop the scrolling.