DI-100 User's Guide

Page: 16 / 19

P/N : 033-02205-Rev2

5.1.4

TARE ©

Function:

Tares, or sets the display ZERO value.

Summary:

When this command is issued, the current A/D converter value is retained as
ZERO output value. All subsequently output values are referenced to this value.

Example:

TARE ←

5.1.5

CAL ©

Function:

Sets (returns) calibration type

Summary:

Sets or returns the currently selected calibration type. Acceptable parameteres
are:

m for milliVolt calibration

2 for 2 point calibration

Example:

CAL m ← (sets mV calibration mode)

5.1.6

mVOLT ©

Function:

Sets (returns) the current mV calibration parameter

Summary:

Sets (returns) the current sensor mV calibration parameter. This may be the
parameter listed on the sensor, the manufacturers data sheet, or may have been
derived through the calibration procedure.

Example:

MVOLT 2.123 ←

5.1.7

2PCAL ©

Function:

Runs the 2 point calibration procedure

Summary:

Begins the 2 point calibration procedure. Follow the prompts from the DI-100 to
complete the calibration as described in the section on calibration above.

Example

2PCAL 250←