Venting the ln, After use – MTS Series 651 Environmental Chambers User Manual

Page 25

Advertising
background image

Series 651 Environmental Chamber

25

Venting the LN

2

after use

After using LN

2

cooling, complete the following steps to prevent coolant from

becoming trapped in the line between the supply tank gas valve and the coolant
solenoid valve.

This procedure must be performed after each test if another test will not be
started within a few minutes.

Do not leave LN

2

coolant in the line after use.

LN

2

left in the line could rupture the line or damage the solenoid when the

liquid warms up and expands.

Always bleed off any LN

2

that may remain in the supply line after testing.

1. When testing is completed, close the liquid withdrawal valve on the supply

tank.

2. Use the controls on the temperature controller to open the solenoid valve.

This allows the coolant in the line to bleed off into the chamber.

CAUTION