How to install a specimen—servohydraulic systems – MTS Advantage Screw Action Grip User Manual

Page 37

Advertising
background image

MTS Advantage™ Screw Action Grips

Operation

37

How to Install a Specimen—Servohydraulic Systems

The following procedure describes the steps for the initial setup and
specimen loading procedures for MTS Advantage Screw Action Grips that
are installed into a servohydraulic test system. This procedure assumes that
the proper set of face surfaces has been selected and properly installed for
the specimen being tested.

Placing hands in the crush zone between the grips while hydraulic
pressure is being applied is dangerous.

This action can result in injury to personnel and damaging forces being
applied to the specimen.

Use extreme care while performing the following steps.

1. Manually adjust the centering guides, located along the side of the

grips, so that the specimen is centered in the grips.

2. Apply full system hydraulic pressure according to applicable system

procedures. Use extreme care not to cause rapid actuator movement
when performing the following steps.

Improper specimen installation can damage the grips or the specimen.

Ensure that the specimen is in full contact with the grip face. To prevent
damage to the grips or specimen, install the specimen precisely as specified
in this procedure.

Ensure that the specimen is not harder than the grip face.

Damage to the grip face can occur if the specimen hardness exceeds H

B

150.

WARNING

CAUTION

CAUTION